Confirming your payment

Skip to content

Sign up now and get .

Competency-Based Questions (CBQ) · Question 4 of 40

Competency: Application A paper mill controls the thickness of its paper with the gauge shown.

Question 43 marksMedium

Competency: Application

A paper mill controls the thickness of its paper with the gauge shown. A krypton-85 source below the moving sheet decays by β⁻ emission (half-life 10.710.7 years), and a Geiger–Müller (G–M) tube above the sheet counts the radiation that gets through. A computer moves the rollers to keep the count rate at a set value.

Class 12 Physics Radioactivity Question 4

With paper of the correct thickness, the count rate (after subtracting the background) is 24002400 per minute. (Hypothetical data.)

(a) Write the equation for the β⁻ decay of krypton-85 (Z=36Z = 36) into rubidium, showing every particle. Explain why an α-source or a γ-source would not be suitable for this gauge.

(b) One morning the count rate rises to 26502650 per minute. Is the paper now too thick or too thin? What must the rollers do?

(c) Show that, because of the source's own decay, the count rate for paper of the correct thickness falls by only about 0.5 %0.5\ \% in 3030 days. (Take ln⁡2=0.693\ln 2 = 0.693, 1 year=3651\ \text{year} = 365 days and e−0.00532=0.9947e^{-0.00532} = 0.9947.)