Competency: Analysis & Evaluation
Each year a hospital physicist checks the X-ray beam of a radiography unit run at kV. Sheets of aluminium are placed one by one across the beam, and a detector behind them measures the intensity. The table gives it as a percentage of the intensity with no aluminium. The thickness of aluminium that halves the intensity is called the half-value thickness. The larger it is, the more penetrating (harder) the beam. The hospital's rule is that at kV the first half-value thickness must be at least mm. (Hypothetical data.)
| Aluminium thickness (mm) | ||||||||
|---|---|---|---|---|---|---|---|---|
| Intensity (%) |
(a) Use the table to find the first half-value thickness, and the extra thickness of aluminium that halves the intensity again (the second half-value thickness).
(b) Use the data to explain why the second half-value thickness is larger than the first.
(c) Does the beam pass the hospital's rule? A trainee suggests that, if a beam ever fails the rule, doubling the tube current will make it more penetrating. Evaluate this suggestion.