Competency-Based Questions (CBQ) · Question 11 of 40

Competency: Analysis & Evaluation Once a year a hospital physicist checks the voltage settings of an X-ray unit.

Question 113 marksMedium

Competency: Analysis & Evaluation

Once a year a hospital physicist checks the voltage settings of an X-ray unit. The tube voltage rises and falls slightly, and its highest value, set on a dial in kilovolts peak (kVp), fixes the shortest wavelength λmin\lambda_{min} of the continuous X-rays. The physicist records the spectrum with an X-ray spectrometer and reads λmin\lambda_{min} from it. The hospital accepts a setting if the true peak voltage is within 5 %5\ \% of the dial value. The readings are given in the table. (Hypothetical data.)

Dial setting80 kVp80\ \text{kVp}60 kVp60\ \text{kVp}
Measured λmin\lambda_{min}0.0161 nm0.0161\ \text{nm}0.0219 nm0.0219\ \text{nm}

Take h=6.63×10−34 J sh = 6.63 \times 10^{-34}\ \text{J s}, c=3.00×108 m s−1c = 3.00 \times 10^{8}\ \text{m s}^{-1} and e=1.60×10−19 Ce = 1.60 \times 10^{-19}\ \text{C}.

(a) Find the true peak voltage at the 80 kVp80\ \text{kVp} setting, and say whether the setting is accepted.

(b) Do the same for the 60 kVp60\ \text{kVp} setting.

(c) For the 110 kVp110\ \text{kVp} setting, find the value of λmin\lambda_{min} that should be read if the dial is exactly right, and the range of readings of λmin\lambda_{min} that can be accepted.