In every range the galvanometer carries Ig=1.0 mA at full scale, and the galvanometer branch and the shunt branch have the same potential difference across them.
10 mA (terminal A): the shunt is the whole chain Rs=R1+R2+R3; the galvanometer branch is G:
IgG=(I−Ig)Rs⇒(0.001)(90)=(0.009)Rs⇒Rs=10 Ω
100 mA (terminal B): the shunt is R2+R3=x; the galvanometer branch is now G+R1=G+(10−x):
(0.001)(100−x)=(0.099)x⇒x=1.0 Ω

1.0 A (terminal C): the shunt is R3=y; the galvanometer branch is G+R1+R2=100−y:
(0.001)(100−y)=(0.999)y⇒y=0.10 Ω
R3=0.10 Ω,R2=1.0−0.10=0.90 Ω,R1=10−1.0=9.0 Ω
Answer: R1=9.0 Ω, R2=0.90 Ω, R3=0.10 Ω. The advantage of this arrangement is that the galvanometer is never left without a shunt while the range is being switched.