(a)
From geometry: G∗=Al=1.501.20=0.800 cm−1
From KCl: G∗=κ×R=1.41×10−3×590=0.832 cm−1
Use 0.832 cm⁻¹. The current does not flow only in the neat block of liquid between the plates; it also spreads around the edges. The plate area and separation are also hard to measure exactly. Calibrating with a solution of accurately known κ captures the real behaviour of the cell.

(b)
κ=RG∗=7560.832=1.10×10−3 S cm−1
Λm=c1000κ=0.005001000×1.10×10−3=220 S cm2 mol−1
(c)
Λm∘=2(50.1)+160.0=260.2 S cm2 mol−1,Λm∘Λm=260.2220=0.85
Na2SO4 is a strong electrolyte and is completely dissociated, so the ratio is not a degree of dissociation. Λm is lower because of interionic attraction: each ion is slowed by its oppositely charged neighbours. The effect is large here because the doubly charged SO42− ion attracts strongly.
Why AC: direct current would electrolyse the solution. Gases or deposits forming on the plates would change their surface and the solution composition, and the measured resistance would drift. A rapidly alternating current reverses before any products build up.